[1]  Specifications are for 60 minute warm–up and select 6½  reading resolution. 
[2]  Relative to calibration standards. 
[3]  20% over range on all ranges, except DCV 1000V, ACV 750V, DCI and ACI 10A range. 
[4]  Specifications are for sine wave input >5% of range. For inputs from 1% to 5% of range and <50 kHz, 
add 0.1% of range additional error. For 50 kHz to 100 kHz, add 0.13% of range.   
750 Vac range limited to 100 kHz or 8x10
7
 Volt-Hz. 
[5]  Specifications are for sine wave input >5% of range. Add 0.1% of the range for the sine wave input is 
1%~5% of the range. 
[6]  For current terminal, > 7A DC or ACRMS for 30 seconds ON and 30 seconds OFF. 
 
 
Low Frequency Performance 
 
Three filter settings are available:   
Slow: 3Hz~300kHz 
Mid:    20Hz~300kHz 
Fast:    200Hz~300kHz 
 
Frequencies greater than these filter settings are specified with no additional errors. 
 
Settling Considerations   
Applying >300VRMS (or >2ARMS) will cause self-heating in signal-conditioning components. These errors are 
included in the instrument specifications. Internal temperature changes due to self-heating may cause additional 
error on lower ac voltage ranges. The additional error will be less than 0.02% of reading and will generally 
dissipate within a few minutes.