Measuring Characteristics 
DC Voltage 
Measurement Method:  Sigma Delta A-to-D converter. 
Input Resistance: 
 200mV, 2V, 20V ranges  Selectable 10MΩ ±  2% or >10GΩ 
 200V, 1000V ranges  10MΩ ±  2% 
 
Resistance 
Measurement Method:  Selectable 4-wire or 2-wire. 
  Current source referenced to LO input. 
Open-circuit Voltage:  Limit in <7V. 
Max. Lead Resistance:  10% of range per lead for 200Ω, 1kΩ range. 1kΩ per lead 
(4-wire ohms)  on all other ranges. 
Input Protection:  1000V on all ranges. 
 
DC Current 
Shunt Resistor:  0.025Ω for 1A, 10A 
1.025Ω for 200mA 
11.025Ω for 2mA, 20mA 
Input Protection:  Externally accessible 10A, 250V fuse 
                                                       Internal 12A, 250V fuse 
 
Continuity / Diode Test 
Response Time:  25 samples / sec   
Measurement Method:    1mA ± 0.2% test current, Limit in <7V 
Continuity Threshold:  Adjustable from 1Ω to 2000Ω 
Input Protection:  1000V 
 
True RMS AC Voltage 
Measurement Method:  AC-coupled True RMS – measures the ac component of input with up to 
400Vdc of bias on any range. 
Input Impedance:  1MΩ ±  2%, in parallel with <100pF 
Input Protection:  750VRMS all ranges 
 
True RMS AC Current 
Measurement Method:  Direct coupled to the fuse and shunt. AC-coupled True RMS measurement 
(measures the ac component only) 
Max. Input:  The DC + AC current peak value <300% of the range. The RMS current 
including DC current <10A.