5-18
Shunt Resistor: 0.025Ω for 1A, 10A,
1.025Ω for 200mA,
11.025Ω for 20mA
Input Protection: Externally accessible 10A, 250V fuse
Internal 12A, 250V fuse
Frequency and Period
Measurement Method: Reciprocal-counting technique. AC-coupled input using the ac voltage
measurement function.
Input Impedance (Voltage Signal): 1MΩ ± 2%, in parallel with <150pF
Shunt Resistor (Current Signal): 0.025Ω for 1A, 10A,
1.025Ω for 200mA,
11.025Ω for 20mA
Input Protection: 750VRMS all ranges;
Externally accessible 10A, 250V fuse
Internal 12A, 250V fuse
Capacitance
Measurement Method: Current input with measurement of resulting ramp.
Connection Type: 2-wire
Triggering and Memory
Samples per Trigger: 1 to 2,000,000
Trigger Delay: 0 to 3600 sec
Input Level: TTL compatible (High level when left trigger input open)
Trigger Condition: Selectable Rising, Falling, Low-level, High-level.
Input Impendence: >20kΩ, in parallel with 400pF, AC-coupled
Delay: <1μs
Min Pulsewidth: 1μs
VMC Output:
Level: TTL compatible (Input to 1kΩ load)
Output Polarity: Selectable Positive, Negative
Output Impendence: 200Ω, typical
Nonvolatile Memory: 512K readings
Volatile Memory: 2M readings

Table of Contents

Other manuals for Rigol DM3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DM3000 Series and is the answer not in the manual?

Rigol DM3000 Series Specifications

General IconGeneral
BrandRigol
ModelDM3000 Series
CategoryMultimeter
LanguageEnglish

Related product manuals