RIGOL
Users Guide for DM3058/DM3058E
6-1
Chapter 6 Characteristics
General Technical Characteristics
Power Supply:
AC 100V~120V, 45Hz~440Hz
AC 200V~240V, 45Hz~66Hz
Power consumption: 20VA peak value
Mechanical:
Height × Width × Depth: 107.0mm×231.6mm ×290.5mm
Weight: 2.5kg
Display:
256×64 LCD display, support dual display, menu, Chinese and English languages,
operation help.
Working Environment:
Precision 0~50, 80% R.H. at 40, No coagulation
Storage Temperate: -20~70
Impact and Shake: According with MIL-T-28800E, III class, 5 Level (Only Sine)
Height above sea level: upper limit 3000m
Safety:
According with IEC61010-1: 2001.
Measure CAT I 1000V, CAT II 600V.
Class of pollution: 2.
Remote Interface:
GPIB (only for DM3058), 10/100Mbit LAN (only for DM3058), USB2.0 Full Speed
Device & Host (support USB flash device), RS232.
Programming Language:
RIGOL DM3058 SCPI, Fluke 45, Agilent 34401A.
LXI Compatibility: LXI Class C, Version 1.1 (only for DM3058).
Warm-up Time: 30min.