[1]  Specifications are for 30 minute warm–up, “slow” measurement rate and 18 ℃~ 28 ℃ 
calibration temperature. 
[2]  20% over range on all ranges, except DCV 1000V, ACV 750V, DCI and ACI 10A range. 
[3]  Specifications are for 4–wire resistance function, or 2–wire resistance using REL operation. 
Without REL operation, add 0.2 Ω additional errors in 2–wire resistance function. 
[4]  For each additional 1 volt over ± 500 VDC add 0.02 mV of error. 
[5]  For current terminal, > 7A DC or AC RMS for 30 seconds ON and 30 seconds OFF. 
[6]  Accuracy specifications are for the voltage measured at the input terminals only. 1 mA test 
current is typical. Variation in the current source will create some variation in the voltage 
drop across a diode junction. 
 
 