EasyManua.ls Logo

Rigol DM3058E

Rigol DM3058E
129 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL
Users Guide for DM3058/DM3058E
6-9
Other Measurement Characteristics
Triggering and Memory
Samples per Trigger: 1 to 2,000
Trigger Delay: 8ms to 2000ms
External trigger input:
Input Level: TTL compatible (High level when left trigger input open)
Trigger Condition: Selectable Rising/Falling edges.
Input Impendence: >20kΩ, in parallel with 400pF, DC-coupled
Min Pulse Width: 500μs
VMC Output:
Electric Level: TTL compensate (Input>=1kΩ load)
Output polarity: positive, negative(selectable)
Output impedance: 200Ω, typical
Arbitrary sensor measurement
Support thermocouple, DC voltage, DC current, resistance (2-wire or 4-wire), capacitance and
frequency output type sensors. Build in thermocouple cold side compensate.
Cold junction compensation: ±3.
Preset ITS-90 transform of B, E, J, K, N, R, S and T thermocouple and transform of platinum Pt100,
Pt385 resistance t temperature sensor.
Math function
Pass/Fail, RELative, Maximum/Minimum/Average, dBm, dB, Hold, Histogram, standard deviation.
History function
Volatile Memory:
2000 reading history record.
Nonvolatile Memory:
10 groups history data storage (2000 readings/group)
10 groups sensor data storage (1000 readings/group)
10 groups instrument settings storage
10 groups sensor settings storage
Support USB flash device external storage extend
w ww. . com
information@itm.com1.800.561.8187

Table of Contents

Related product manuals