3.  Choose a proper Voltage range according to the measured circuit. 
 
Table 2-1 Measurement Characteristics of DC Voltage 
Ranges*  200mV, 2V, 20V, 200V, 1000V 
Input Protection  1000V on all ranges (HI terminal) 
Parameters 
Range, DC input impedance, REL 
NOTE*: All the ranges enable to obtain 20% value higher than original except 1000V. 
Besides, both Manual and Auto are available for setting every range. When inputting range 
is higher than 1000V at 1000V Level, “OVER LOAD” will be shown on the screen. 1000V 
input protection exists in every range. 
 
4.  Set the DC input impedance 
Press   Res to set the DC resistance as “10 MΩ (default value)” or “>10 
GΩ”. Users can execute DC voltage measurement directly without modifying 
this parameter which has been setup before leaving factory. 
 
5.  Set relative value (Optional) 
 Press REL to open or close Relative math function. When it is open, “REL” is 
shown over the screen and the reading displayed is a relative value which 
comes from the result of actual measurement value subtracts the value that 
has been set. (For the details about how to set, please refer to “
Math 
Functions” in Chapter 2.) 
 
6.  Read measurement result  
Select required measurement (reading) rate by using Left or Right direction 
keys and read the measurement result. 
 
7.  View history data 
Press History to enter into the following interface and check the data getting 
from this measurement and saved.  
 
Figure 2-6 View interface of history data 
 
w ww. . com
information@itm.com1.800.561.8187