Short-circuit Resistance
Set up the short-circuit resistance value in the short-circuit test menu. When the
measured resistance is lower than the short-circuit resistance, the circuit is
considered as connected, and the beeper sounds (if sound is on). The
short-circuit resistance is only applicable to the continuity test.
Press Conti, enter the menu shown below:
Figure 2-59 Set up the short-circuit resistance
Use direction keys to change the parameter values:
Press left and right directional keys to choose different digits. Press up and down
keys to change the current digit value.
Continue Resistance
The range of short-circuit resistance is 1Ω~2000Ω. The default value is
10Ω.
The short-circuit resistance value is stored in the nonvolatile memory and
the resistance remains unchanged after power-off.