10
10. Power Fuse
The specification of the fuse of M300 is AC 250V T3.15 A. If a new fuse is
required, please refer to the following steps.
1) Turn off the instrument and remove the power cord.
2) Insert the small straight screw driver into the slot at the power socket and
prize out the fuse seat.
3) Replace a specified fuse.
4) Insert the fuse seat into the slot.
CAUTION
Please use the specified power fuse and make sure that the fuse
seat is not short-circuited to avoid electric shock and fire.