Module Overview
1. MC3065 (DMM Module):
The DMM module is used to measure the signals under test and has 6½ digits
reading resolution. The measurement functions include DCV, ACV, 2-wire
resistance, 4-wire resistance, frequency, period, temperature and any sensor.
Note: After connecting the DMM module, make sure that the signal under test
connected to the analog bus is no greater than 300 Vdc or 300 Vrms.
2. MC3132 (32-Channel Multiplexer):
32-channel multiplexer.
All 32 channels switch both HI and LO inputs, thus providing fully isolated inputs
to the DMM module. MC3132 is divided into two banks (called A and B) of 16
two-wire channels each. When making 4-wire resistance measurement, the
instrument automatically pairs channels of A bank and B bank. All channels are
break-before-make. You can close multiple channels on this module only if you
have not configured any channel to be part of the scan list. This module can be
connected with MC3065 (DMM Module, if MC3065 is currently inserted). For the
specifications of this module, please refer to the User’s Guide or Data Sheet of
this product.
3. MC3164 (64-Channel Single-Ended Multiplexer):
64-channel single-ended multiplexer.
All 64 channels can switch HI input only. MC3164 is divided into two banks (A
and B) of 32 single-ended channels each. All channels are break-before-make.
In any case, you cannot close multiple channels on this module. This module can
be connected with MC3065 (DMM Module, if MC3065 is currently inserted) but
cannot be used for 4-wire resistance measurement. For the specifications of this
module, please refer to the User’s Guide or Data Sheet of this product.
4. MC3232 (32-Channel Reed Multiplexer):
32-channel reed multiplexer.
MC3232 is applicable to high-speed scan and high-throughput auto test
applications. All 32 channels can switch both HI and LO inputs, thus providing
fully isolated inputs to the DMM module. MC3232 is divided into two banks (A
and B) of 16 two-wire channels. When making 4-wire resistance measurement,
the instrument automatically pairs channels of A bank and B bank. All channels
are break-before-make. You can close multiple channels on this module only if
you have not configured any channel to be part of the scan list. This module can
be connected with MC3065 (DMM Module, if MC3065 is currently inserted). For
the specifications of this module, please refer to the User’s Guide or Data Sheet