4. Alarm Functions 4-4. Outside operating temperature range warning
40 / 79
4-4. Outside operating temperature range warning
An outside operating temperature range warning (temperature range error) will be issued if a product (other than the
OX-04G) is used for 20 minutes or more outside the operating temperature range.
When a temperature range error occurs, either leave the product for five minutes or longer in the operating temperature
range, or turn off the power of the main unit.
If an outside operating temperature range warning occurs, the user will be alerted by the audible buzzer and flashing
alarm lamp.
Alarm type Outside operating temperature range warning
Buzzer Repeated intermittent beeps at about 1-second intervals: “Beep”
Alarm lamp Repeated flashing at about 1-second intervals
LCD display
Display example: Outside operating temperature range warning
NOTE
Press the POWER/MODE button to reset the alarm.
The outside operating temperature range warning does not apply to the OX-04G.