7. Maintenance 7-1. Maintenance intervals and maintenance items
67 / 79
7
Maintenance
The product is an important safety and disaster-prevention device.
Perform product maintenance at regular intervals to ensure performance and to improve disaster-prevention and safety
reliability.
7-1. Maintenance intervals and maintenance items
Maintain the following items at regular intervals:
Daily maintenance: Perform maintenance before commencing work.
Monthly maintenance: Perform alarm tests monthly. (Refer to ‘7-4. Performing alarm tests’ in the Technical Manual.)
Regular maintenance: Perform maintenance at least once a year (ideally, at least once every six months).
Maintenance
item
Maintenance details
Daily
maintenance
Monthly
maintenance
Regular
maintenance
Battery level
Check to confirm that battery levels are adequate.
Concentration
display
Check to confirm that the concentration readout is [0 ppm] ([20.9 %]
for oxygen) by measuring fresh air.
If the readout is not [0 ppm] ([20.9%] for oxygen), check to confirm
that no interference gases are present, then perform air calibration.
Main unit
operation
Check to confirm that no fault alarm is displayed on the LCD display.

Other manuals for Riken Keiki 04 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Riken Keiki 04 Series and is the answer not in the manual?

Riken Keiki 04 Series Specifications

General IconGeneral
BrandRiken Keiki
Model04 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals