HS-04 Operator’s Manual Chapter 4: Display Mode • 29
Chapter 4: Display Mode
This section describes Display Mode which is accessible from Measuring Mode. See Table 7
below for a list of Display Mode’s menu items, a short description of each item, and the page
number for further description.
Tips for Using Display Mode
• To enter Display Mode and scroll from one item to the next or skip an item when a question is
asked, press and release POWER MODE.
• To enter an item, press and release AIR.
• To change a flashing parameter, press and release AIR. To reverse the movement in a list (ie.
from down to up or vice versa):
a. Press and hold AIR.
b. Immediately press POWER MODE and then release both buttons.
Table 7: Display Mode Menu Items
Display Mode Menu
Item
Description
PEAK (page 30) Displays the Peak reading.
STEL (page 30) Displays the STEL reading.
TWA (page 31) Displays the TWA reading.
USER ID (page 31)
A
View and/or change the User ID.
STN ID (page 32)
A
View and/or change the Station ID.
CAL.DATA (page 33)
B
Displays the last calibration date.
BP.DATA (page 33)
C
Displays the last bump test date.
DATE (page 34) Displays the current date and time.
TEMP (page 34) Displays the current temperature.
ALARM--P (page 34) View alarm points.
BUZZ.VOL (page 35)
D
Set the buzzer volume to LO or HI (factory setting).
A
Only appears if DISP.SET is set to ON in User Mode (factory setting) and if ID DISP is set to ON in
Maintenance Mode (factory setting is OFF).
B
Only appears if CAL.RMDR is set to ON in User Mode (factory setting).
C
Only appears if BP.RMDR is set to ON in User Mode (factory setting is OFF).
D
Only appears if DISP.SET is set to ON in User Mode (factory setting).