66 • Chapter 5: User Mode and Calibration HS-04 Operator’s Manual
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the KEY.TONE item in
User Mode.
5. See “Entering Measuring Mode (START)” on page 69 to enter Measuring Mode.
Display Mode Items (DISP.SET)
OFF: USER ID, STN ID, and BUZZ.VOL screens do not appear in Display Mode.
ON (factory setting): BUZZ.VOL screen appears in Display Mode. USER ID and STN ID
screens appear in Display Mode if ID DISP in Maintenance Mode is also set to ON (factory
setting is OFF).
1. While in User Mode, press AIR to scroll to DISP.SET.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the DISP.SET item in User
Mode.
5. See “Entering Measuring Mode (START)” on page 69 to enter Measuring Mode.
Zero Suppression (ZERO.SUP)
This item only appears if ZSUP.DSP is set to ON in Maintenance Mode (factory setting is OFF).
The ZERO.SUP setting is not intended for field adjustment. The default setting is ON. The
suppression value is 0.3 ppm.
Zero Follower (ZERO.FLW)
This item only appears if ZFLW.DSP is set to ON in Maintenance Mode (factory setting is OFF).
The ZERO.FLW setting is not intended for field adjustment. The default setting is ON.