56Chapter 5: User Mode and Calibration HS-04 Operator’s Manual
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the CAL.EXPD item in the
CAL SET menu.
5. See “Exiting the CAL SET Menu” on page 56 to return to User Mode.
Exiting the CAL SET Menu
1. While in the CAL SET menu, press AIR to scroll to ESCAPE.
2. Press and release POWER MODE to return to the CAL SET item in User Mode.
3. See “Entering Measuring Mode (START)” on page 69 to enter Measuring Mode.
Setting Bump Test Parameters (BUMP.SET)
The BUMP.SET menu has 5 items: SETTING, BP.RMDR, BP.INT, BP.EXPD, and ESCAPE.
1. While in User Mode, press AIR to scroll to BUMP.SET.
2. Press and release POWER MODE. The SETTING item appears.
SETTING
The SETTING menu has 5 items: GAS.TIME, CHECK, CAL.TIME, A-CAL, and ESCAPE.
1. After entering the BUMP.SET menu, press AIR to scroll to SETTING.
2. Press and release POWER MODE. The GAS.TIME item appears.
BUMP .SET
SETTING
SETTING