54 • Chapter 5: User Mode and Calibration HS-04 Operator’s Manual
10. Use AIR to scroll to ESCAPE.
11. Press and release POWER MODE to return to the CAL--P item in the A-CAL menu.
12. Press AIR to scroll to ESCAPE.
13. Press and release POWER MODE to return to the A-CAL item in the GAS CAL menu.
14. See “Exiting the GAS CAL Menu” on page 54 to return to User Mode.
Exiting the GAS CAL Menu
1. While in the GAS CAL menu, press AIR to scroll to ESCAPE.
2. Press and release POWER MODE to return to the GAS CAL item in User Mode.
3. See “Entering Measuring Mode (START)” on page 69 to enter Measuring Mode.
Setting Calibration Parameters (CAL SET)
The CAL SET menu has 4 items: CAL.RMDR, CAL.INT, CAL.EXPD, and ESCAPE.
1. While in User Mode, press AIR to scroll to CAL SET.
2. Press and release POWER MODE. The CAL.RMDR item appears.
CAL.RMDR
ON (factory setting): The HS-04 gives an indicator at start up if it is due for calibration. The type
of indicator depends on the CAL.EXPD setting (see page 55).
OFF: The HS-04 does not give an indicator at start up if it is due for calibration.
1. After entering the CAL SET menu, press AIR to scroll to CAL.RMDR.
2. Press and release POWER MODE. The current setting flashes.