HS-04 Operator’s Manual Chapter 5: User Mode and Calibration53
Setting the Calibration Values in CAL--P
1. While in User Mode, press AIR to scroll to GAS CAL.
2. Press and release POWER MODE. The AIR item appears.
3. Use AIR to scroll to the A-CAL item (E-CAL if E-CAL in User Mode is set to ON).
4. Press and release POWER MODE. The Auto Cal screen displays.
5. Use AIR to scroll to CAL--P.
6. Press and release POWER MODE. The gas name displays.
7. Press and release POWER MODE. The calibration value begins to flash.
8. Use AIR to adjust the calibration gas value. The calibration gas value in the instrument must
match the value listed on the calibration gas cylinder you are using.
9. Press and release POWER MODE to save the change. The calibration gas value stops flashing
and the unit returns to the gas name screen.
GAS CAL
A-CAL
H2S
50.0
A-CAL
ppm
CAL-P
H2S
CAL-P
ppm
H2S
50.0
CAL-P
ppm