44 • Chapter 5: User Mode and Calibration HS-04 Operator’s Manual
Option C from Flow Chart Option D from Flow Chart
•Bump test failed
• A-CAL set to ON (factory setting)
• Calibration passed
• Bump test failed
• A-CAL set to ON (factory setting)
• Calibration failed
1. A calibration immediately and
automatically starts. Continue to apply
the calibration gas.
2. The calibration time is the difference
between the GAS.TIME and the
CAL.TIME values defined in the
BUMP.SET\SETTING item in User
Mode.
3. The instrument shows the pass/fail
results of the bump test/calibration.
Use AIR to scroll between the results
and the readings.
4. Close the regulator.
5. Unscrew the regulator.
6. Remove the calibration cup.
7. Press and release POWER MODE to
return to the BUMP screen in the
BUMP menu.
8. Use AIR to scroll to START and press
and release POWER MODE to enter
Measuring Mode.
1. A calibration immediately and
automatically starts. Continue to apply
the calibration gas.
2. The calibration time is the difference
between the GAS.TIME and the
CAL.TIME values defined in the
BUMP.SET\SETTING item in User
Mode.
3. The instrument shows the pass/fail
results of the bump test/calibration.
The LED flashes and the buzzer
sounds. Use AIR to scroll between the
results and the readings.
4. Close the regulator.
5. Unscrew the regulator.
6. Remove the calibration cup.
7. Press and release POWER MODE to
return to the BUMP screen in the
BUMP menu.
8. Use AIR to scroll to START and press
and release POWER MODE to enter
Measuring Mode.
H2S
0.0
CAL 60
ppm
H2S
APLY 60
ppm
0.0
H2S
F P
BMP / CAL
ppm
H2S
15.0
BMP /
ppm
H2S
24.0
/ CAL
ppm
AIR
AIR
H2S
CAL 60
ppm
H2S
APLY 60
ppm
0.0 0.0
H2S
F F
BMP / CAL
ppm
H2S
0.0
BMP /
ppm
H2S
5.0
/ CAL
ppm
AIR
AIR