64Chapter 5: User Mode and Calibration HS-04 Operator’s Manual
BEEP.SEL
BEEP.SEL defines what kind of confirmation or non-compliance indicator you want to occur in
Measuring Mode. The available choices are:
OFF (factory setting): The HS-04 does not provide a confirmation alert or non-compliance
indicator.
LED: The HS-04’s LED double flashes as often as defined by the BEEP.INT parameter to verify
that the instrument is operating.
BUZZER: The HS-04’s buzzer double beeps as often as defined by the BEEP.INT parameter to
verify that the instrument is operating.
LED+BUZ: The HS-04’s LED double flashes and the buzzer double beeps as often as defined by
the BEEP.INT parameter to verify that the instrument is operating.
BMP/CAL: If a bump test or a calibration is due and if BP.EXPD or CAL.EXPD is set to
CONFIRM (factory setting) or NONE, the HS-04’s LED double flashes as often as defined by
the BEEP.INT parameter to indicate a non-compliance. Once a bump test or calibration
(depending on which is due) is done, the LED stop flashing.
1. While in the BEEP menu, press AIR to scroll to BEEP.SEL.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the BEEP.SEL item in the
BEEP menu.
5. See “ESCAPE” on page 65 to return to User Mode.
BEEP.INT
The BEEP.INT parameter defines how often the confirmation alert or non-compliance indicator
selected in BEEP.SEL occurs. This setting only applies if the BEEP.SEL parameter is set to
something other than OFF (factory setting). The available choices are 0.5 minutes and 1-99
minutes in 1 minute increments. The factory setting is 5 minutes.
1. While in the BEEP menu, press AIR to scroll to BEEP.INT.
2. Press and release POWER MODE. The current setting flashes.
BEEP .SEL
BEEP .INT

Table of Contents