8. Install the new sensor in the socket on the rear case. The colored side faces away from the rear
case and the slots in the sensor line up with tabs in the socket.
CAUTION: Forcing a sensor into a socket without lining the sensor slots up with the socket tabs
may damage the sensor or socket.
9. Reinstall the front case to the rear case.
10. Reinstall the 4 screws that were removed in Step 3.
11. Reinstall the rubber boot, if being used.
12. Calibrate the H
2
S response as described on page 43.