R&S FSP Module Replacement
1164.4556.82 3.21 E-1
Replacing Hard Disk A60
(See chapter 5, spare parts list, item 710, and explosion drawings 1164.4391 and 1164.4404.)
The hard disk is incorporated between the front module and the analog boards.
The spare part is already formatted for the R&S FSP and contains the complete software.
Opening the Instrument and Removing the Hard Disk
Switch off the instrument, pull the mains plug, unscrew the 4 rear-panel feet (450) and push the tube
(410) backwards and take off.
Unscrew the two countersunk screws (260) at the top of the instrument and the three countersunk
screws (250) at the fan side, lift the cover at the top (240) of the instrument to the left side and take
off.
Disconnect the SATA data cable (721) and the SATA power cable (722) at the hard disk.
Unscrew the two countersunk screws (740) on the hard disk mounting plate (730).
Take the hard disk (710) completely with the hard disk mounting plate (730 out of the R&S FSP.
Unscrew the four countersunk screws (750) remove the hard disk and mount a new hard disk to the
mounting plate (730).
Installing the New Hard Disk and Putting into Operation
Fit the hard disk and the disk holder with two screws (740) to the instrument
Note: Be careful with the lower cover plate locking in the corresponding opening.
Connect the SATA data cable (721) and the SATA power cable (722) to the hard disk.
Note: Connect cable as drawing 1164.4404.
Put the top cover (240) on the instrument and mount it with two countersunk screws (260) and three
countersunk screws (250).
Push the tube (410) on the instrument and mount the 4 rear-panel feet (450).
Connect the instrument to the mains and switch on the instrument power switch. The instrument is
now in stand-by mode.
After the change of the hard disk a cold boot is necessary. See section "Performing a Cold Boot" in
this chapter.
After starting the instrument check for system messages:
- [ SETUP : SYSTEM INFO : SYSTEM MESSAGES ]
Start selftest and check results:
- [ SETUP : SERVICE : SELFTEST ], then [SELFTEST RESULT]
Start a total calibration and check the result:
- [ CAL : CAL TOTAL ] , then [ CAL RESULTS ]
Perform a firmware update if necessary.
Repeat the total calibration and check the result:
- [ CAL : CAL TOTAL ] , then [ CAL RESULTS ]