R&S FSP  Performance Test 
1164.4554.82 1.29 E-1 
Checking Level Accuracy and Frequency Response with 
Preamplifier   
Test equipment: 
- Signal generator : 
R&S FSP 3:          Refer to "Measurement Equipment", item 3. 
R&S FSP 7:          Refer to "Measurement Equipment", items 3 & 5.
Frequency range: 
  R&S FSP 3:  10 MHz to 3 GHz 
  R&S FSP 7:  10 MHz to 7 GHz 
 Maximum level   0 dBm 
- Power meter (refer to "Measurement Equipment", item 8) 
- Power sensor: 
 R&S FSP 3:   Refer to "Measurement Equipment", item 9. 
 R&S FSP 7:   Refer to "Measurement Equipment", items 9 & 10. 
 
Frequency range: 
  R&S FSP 3:  10 MHz to 3 GHz 
  R&S FSP 7:  10 MHz to 7 GHz 
  maximum power   P
max
  100 UW
RSS referenced to indicated power 
1 MHz to 1 GHz    1.5 % 
1 GHz to 7 GHz   2 %
impedance Z = 50 
- Power splitter (refer to "Measurement Equipment", item 6) 
 
Frequency range: 
  R&S FSP 3:  10 MHz to 3 GHz 
  R&S FSP 7:  10 MHz to 7 GHz 
 Level imbalance
1
) 1 MHz to 1 GHz   0.1 dB 
  1 GHz to 7 GHz   0.2 dB 
 
1
) If a power divider with higher level imbalance is used, correction 
  of the measured frequency response is recommended.