R&S FSP Measurement Equipment
1164.4556.82 1.1 E-1
1 Performance Test
Test Instructions
The rated specifications of the analyzer are tested after a warm-up time of at least 15 minutes and
overall calibration. Only in this case can compliance with the specified data be ensured. Start of
overall adjustment: [CAL : CAL TOTAL]
Unless specified otherwise, all measurements will be performed with external reference frequency.
Values specified in the following sections are not ensured. Only the technical specifications of the
data sheet are binding.
The values specified in the data sheet are the ensured limits. Due to measurement errors, these
limits must be extended by the tolerance of the measurement equipment used in this performance
test.
Inputs for settings during measurements are shown as follows:
[<KEY>] Press a key on the front panel, e.g. [ SPAN ] .
[<SOFTKEY>] Press a softkey, e.g. [ MARKER -> PEAK ].
[<nn unit>] Enter a value and terminate by entering the unit, e.g. [ 12 kHz ]
Successive entries are separated by [:], e.g. [ BW : RES BW MANUAL : 3 kHz ]
Measurement Equipment and Accessories
Item Type of
equipment
Specifications recommended Equipment
recommended
Alternative
equipment
Use
1 Frequency
counter
Accuracy < 1x10
-9
,
Frequency range up to 10 MHz
Advantest
R5361B
with option 23
HP 53132A
with option
012 (Agilent)
Frequency accuracy of
reference oscillator
2 Signal generator R&S FSP 3: 10 MHz to 10 GHz
R&S FSP 7: 10 MHz to 10 GHz
R&S FSP 13: 10 MHz to 13.6 GHz
R&S FSP 30: 10 MHz to 30 GHz
R&S FSP 40: 1 MHz to 40 GHz
R&S SMP02
R&S SMP02
R&S SMP02
R&S SMP04
R&S SMP04
HP 63620 B
HP 63620 B
HP 63620 B
HP 83640 B
HP 63640 B
(Agilent)
Immunity to interference
Third order intercept
Frequency response
3 Signal generator 1 MHz to 3 GHz
Phase noise at 498 MHz:
< -100 dBc/Hz @ 100 Hz
< -115 dBc/Hz @ 1 kHz
< -127 dBc/Hz @ 10 kHz
< -130 dBc/Hz @ 100 kHz
< -142 dBc/Hz @ 1 MHz
R&S SMA100A
with options
R&S
SMA-B103,
R&S SMA-B22
HP 8664 A
OPT 004
(Agilent)
Calibration source
128 MHz
2nd order harmonic dist.
Third order intercept
IF filters
Frequency response
Display linearity
RF attenuator
Reference level switching
Phase noise
RF trigger
4 Signal generator 498 MHz
Phase noise at 498 MHz:
< -130 dBc/Hz @ 100 kHz
< -142 dBc/Hz @ 1 MHz
HP 8663 A
(Agilent)
Phase noise (only if used
with HP 8664 A)