EasyManua.ls Logo

Rohde & Schwarz FSP7 - Checking Noise Display

Rohde & Schwarz FSP7
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Test R&S FSP
1164.4556.82 1.10 E-1
Checking Noise Display
Test equipment:
50 termination (refer to "Measurement Equipment", item 7)
Frequency range R&S FSP 3: to 3 GHz
R&S FSP 7: to 7 GHz
R&S FSP 13: to 13.6 GHz
R&S FSP 30: to 30 GHz
R&S FSP 40: to 40 GHz
Test setup:
Terminate the RF input of the R&S FSP with 50 .
R&S FSP settings:
- [ PRESET ]
- [ AMPT : RF ATTEN MANUAL : 0 dB ]
- [ SPAN : 0 Hz ]
- [ BW: RES BW MANUAL : 10 Hz ]
- [ BW : VIDEO BW MANUAL : 1 Hz ]
- [ BW : SWEEP TIME MANUAL : 0.1 s ]
- [ TRACE 1 : AVERAGE ]
- [ TRACE 1 : SWEEP COUNT : 30 ENTER ]
- [ AMPT : {RefLev} ]
- [ FREQ : CENTER : {f
n
} ]
Refer to table below for values of RefLev.
Refer to "Performance Test Report" table for values of f
n
.
Measurement:
Set marker to peak:
- [ MKR : PEAK ]
Evaluation:
The noise level is displayed by the level reading of marker 1.
Frequency
< 10 kHz < 100 kHz < 1 MHz < 10 MHz > 10 kHz
RefLev
-10 dBm -20 dBm -30 dBm -60 dBm -60 dBm

Table of Contents

Related product manuals