Performance Test R&S FSP
1164.4556.82 1.22 E-1
Performance Test Option Tracking Generator – R&S FSP-B9
Checking Output Level
Test equipment:
- N cable (refer to "Measurement Equipment", item 14)
Frequency up to 3 GHz
Maximum attenuation < 0.2 dB
Test setup:
Connect tracking generator output to RF input of the R&S FSP.
R&S FSP settings:
- [ PRESET ]
- [ MODE NETWORK ]
- [ FREQ : CENTER : 128 MHz ]
- [ SPAN : 0 Hz ]
- [ BW : RES BW MANUAL : 1 kHz ]
- [ AMPT : RF ATTEN MANUAL : 10 dB ]
- [ AMPT : -0 dBm ]
- [ NETWORK : SOURCE POWER {level}]
{level} : 0 dBm; -5 dBm; -10 dBm; -15 dBm; -20 dBm; -25 dBm
Measurement:
Set marker to peak:
- [ MKR : PEAK ]

Table of Contents