R&S FSP Troubleshooting
1164.4556.82 3.69 E-1
Troubleshooting
Malfunctions may have simple causes but also may be caused by faulty components.
These troubleshooting instructions allow for locating the error causes down to board level and make the
instrument ready for use again by means of board replacement.
We recommend that the instrument be shipped to our experts in the service centers (cf. address list) for
module replacement and further error elimination.
WARNING
Do not plug or unplug boards prior to disconnecting them from ac supply!
Be careful not to cause short-circuits when measuring voltages!
The following utilities are provided in the R&S FSP for diagnostic purposes:
Permanent monitoring of levels and frequencies in the instrument
Selftest
System error correction
Note: When problems occur, first check, whether any connections (cables, plug-in connections of
boards etc.) are damaged or wrongly connected.
Measuring Equipment and Accessories
Item Type of equipment Specifications recommended Equipment
recommended
R&S-
Order No.
Use
1 DC meter R&S URE 0350.5315.02 Troubleshooting
2 Spectrum analyzer Frequency range 0 to 7 GHz R&S FSEB 20 1066.3010.20 Troubleshooting
3 Adapting cable 1 m long
R&S SMP- to SMA connection
- 1129.8259.00 Troubleshooting
4 Adapting cable 0.5 m long
SMP-to SMP-connection
- 1129.8265.00 Troubleshooting
5 Adapter board Extension 150 mm high
48-contact, 2 mm spacing
- 1100.3542.02 Troubleshooting