R&S SMA100A Information on the Current Version and History
Release Notes for Firmware 1171.5360.52 10
1.8 Version 2.20.470.30
Released: June 2013
Fixed Issues
● DME Modulation (R&S® SMA-K26)
– The selection of the ICAO channels was too long.
– The parameter Range Delay was wrong, if ICAO channels were selected.
– In rare cases, the entered squitter rate was not accepted, if it was entered
using the front panel keys.
– In rare cases, the parameter Range Distance was rounded in a wrong way.
● In rare cases, the error message ‘CWOP6 PLL unlocked’ appeared, if the
frequency was changed from a value below 3 GHz to a value above 3 GHz.
● Phase Modulation (R&S® SMA-B20 or R&S® SMA-B22): The parameter PhiM Int
Deviation could not be completed by the *1/Enter key.
● LF Generator: The SCPI command [:SOURce]:LFOutput<ch>:INTernal:VOLTage
does not work for <ch> greater than 1.
1.9 Version 2.20.470.18
Released: August 2012
New Functionality
● NRP-Z Level Control: Level can be continuously controlled using NRP-Z power
sensors.
● New Setting “Frequency Multiplier” enables the instrument to modify the displayed
frequency value by a user definable factor.
● Support of IVI-6.1 High Speed LAN instrument protocol (HiSLIP)
● LAN Services can be enabled and disabled individually to prevent unintended
access to instrument
● Frequency and Level displays can be annotated in SETUP Security in order to
conceal the current frequency and level and showing only asterisks instead.
● New remote control emulation R&S® SMT03
● New remote control emulation R&S® SMY01 and R&S® SMY02
Modified Functionality
● VOR Modulation (R&S® SMA-K25): Added ICAO channels, if Carrier Freq. Knob
Step is set to Defined.