R&S SMA100A Information on the Current Version and History
Release Notes for Firmware 1171.5360.52 17
1.18 Version 2.05.220.13
Released: December 2008
Fixed Issues
File manger shows invalid directories
Bugfix for selftest
1.19 Version 2.05.220.10
Released: November 2008
Modified Functionality
Remote control emulation: The selected remote control emulation will be displayed
in the info line
Fixed Issues
The button “Adjust DME Pulse Slope” in the dialog “Internal Adjustemnts” does not
have any function until the DME dialog would be opened for the first time
During the DME analysis the error message “Err-222 Value out of range” appears
SMA-B103L and SMA-B106L does not support DME
The tooltip does not display the correct values and units
Changes at the level units does not work
The BUSY indicator does not appear during the whole preset time
Some read only parameters are not marked as read only parameters
Downgrade to a firmware version 2.05.68.18 or lower does not work
1.20 Version 2.05.220.02
Released: October 2008
New Functionality
New feature DME Modulation (Option R&S® SMA-K26)
New feature Power Analysis (Option R&S® SMA-K28)
New feature remote control emulation HP 8642
New feature remote control emulation HP 8645
New feature remote control emulation HP 8647/8648