R&S SMA100A Information on the Current Version and History
Release Notes for Firmware 1171.5360.52 20
● New function “Factory Preset” (Setup-Menu) resets settings not covered by
PRESET Key
● PRESET (*RST) does not affect reference oscillator settings any more
Fixed Issues
● Remote Control: The command SOURce:POWer UP does not work
● Modulation ILS-GS/ILS-LOC: The DDM could not be entered in µA
● Remote Control Emulation: A user defined selection of *IDN? does not have any
effect
● Remote Control Emulation: *IDN? selection does not saved while firmware
shutdown
● Remote Control Emulation Aeroflex203x/4x: CFRQ? Does not return the current
frequency
● Remote Control: Time parameters do not work in exponential description
● Listmode: Listmode does not work for f > 3GHz, if SMA-B106 and SMA-B20
installed
● Modulation ILS-GS/ILS-LOC: No coupling between GS and LOC
● Modulation ILS-LOC: Setting of modulation frequency does not have any effect
● MenuSetupEthernet does not open the remote settings for Ethernet
● Online Help: Missing help pages for few pulse generator and listmode parameters
● Save/Recall: Could not recall saved setups
1.24 Version 2.05.04
Released: November 2007
New Functionality
● New feature remote control emulation for Aeroflex
2030/2031/2032/2040/2041/2042
● New feature remote control emulation for Racal 3102
Modified Functionality
● Support of new interface board (IfBoard) with part number 1400.0530.02
● Sweep: Minimum dwell time change from 10 ms down to 3 ms
● Pulse generator: New limits for pulse period (20 ns to 100 s), pulse width (5 ns to
100 s), pulse Delay (10 ns to 100 s), double pulse width (5 ns to 100 s), double
pulse delay (10 ns to 100 s) and resolution (5 ns)