Revised 10/02/2012 1.13
Storage:
Temperature Range:-10°C to +65°C
Relative Humidity: 30% to 95%, non-condensing
Relative Atmospheric Pressure: 70 to 106 kPa (525 to
795 mmHg)
Dimensions:
Height: 162 cm (63.8 in.)
Width: 170 cm (66.9 in.)
Depth: 70 cm (27.6 in.)
Weight: 465 kg (1023 lbs)
Instrument Life Expectancy: 7 years (with
instrument powered on 24 hours a day, every day)
Safety Standards
Electrical Safety:
For Model 7720 –
cETL Certification
Complies with:
UL 61010-1: 2
nd
Ed.: 2004,
CAN/CSA-C22.2 No. 61010-1: 2
nd
Ed.: 2004,
CAN/CSA-C22.2 No. 61010-2-010:04,
IEC61010-2-010: 2003
CAN/CSA-C22.2 No. 61010-2-081:04,
IEC61010-2-081: 2001
CAN/CSA-C22.2 No. 61010-2-101: 04,
IEC61010-2-101: 2002
FDA Registration (Regulation No. 864.3875)
For Model 7721 –
Japanese Pharmaceutical Law
Complies with:
Quality Management System- ISO 13485: 2003
Electrical Safety
JIS C 1010-1: 2005
IEC61010: 2001 (ed2)
IEC61010-2: 010: 2003
IEC61010-2-081: 2001
IEC61010-2-101: 2002
EMC- JIS C1806-1: 2001
Risk Management- JIS T 14971: 2003
For Model 7722 –
Complies with:
LVD, IVD
EN61010-1: 2001 (ed2)
EN61010-2-010: 2003 (ed2)
EN61010-2-081: 2001
EN61010-2-101: 2002
EMC
CISPR11: 1999 Group 1 Class A (EN55011)
EN50082-1: 1997
EN61000-3-3: 1994 Amendment 1&2 (2001)
EN61000-4-2: 1995 Amendment 1&2 (2001)
EN61000-4-3: 2002
EN61000-4-4: 2004
EN61000-4-5: 1995 Amendment 1
EN61000-4-6: 1996
Amendement1&2&3&correction1
EN61000-4-11: 2004
Regulatory Information:
FDA Certified with ETL evaluation (USA); complies
with requirements of The Pharmaceutical Affaires
Law (Japan); CE Certified in compliance with the re-
quirements of EMC Directive, LV Directive in Europe
(Europe)