CUSTOMIZATION OF SETTINGS
3.2 Revised 10/02/2012
SOFTWARE VERSION
displays info
r-
mation related to instrument software and
firmware versions and serial n
umber (see
Viewing the Software Version
on page
4.14 for details).
SYSTEM SETUP allows for the config
u-
ration of “user-se
lectable” options and
operating parameters (see “
Configuring
System Settings” on page 3.2 for details).
PASSWORDS allows for th
e addition,
modification,
and removal of system users
(see “Setting Names and Passwords
on
page 3.4 for details).
USER DEFINED FILE
allows for the
addition, modification,
and removal of
assignable “identifiers” for use as sorting
criteria for the Run History File and
Statistics (see “Managing the User D
e-
fined File” on page 3.6 for details).
READY / SHUTDOWN TIMES esta
b-
lishes the start-up and shut-
down times
for automatic operation of the system (see
Setting Ready / Shutdown Times
” on
page 3.7 for details).
PROCESS RECOVERY
allows for
initia
tion of a “recovery process” for
baskets in Retorts #3 and #4 when pr
o-
cessing is interrupted (aborted) due to an
error condition (see “
Section 5, Recovery
Process” for details).
NOTE: The
PROCESS RECOVERY
button is enabled only during the
WARMING UP SYSTEM cycle.
Configuring System Settings
The SYSTEM SETUP option allows changes to “user-
selectable” and “user-specified” system settings and
parameters.
To change system settings:
1. From the UTILITIES screen, touch the SYSTEM
SETUP button.
SYSTEM SETUP screen, page 1, displays (Figure
3-B), showing the currently selected parameters
for the instrument (selected parameters are indi-
cated by a “gray” background).
Figure 3-B
NOTE: The SYSTEM SETUP screen is a multi-page
display screen.
Use the PREVIOUS and NEXT buttons in the
lower right-hand corner of the screen to toggle
between pages.
2. Make changes to the following system settings
/parameters on SYSTEM SETUP, page 1, as desired:
REAGENT CONTROL sets the units by which
reagent use is calculated (a cumulative count of
the selected units is shown on the WARMING UP
SYSTEM screen, MANUAL START screen and
AUTO START screen).
Touch the desired option field, either
CASSETTES or RUN.
The factory default setting for REAGENT
CONTROL is CASSETTES.
NOTE: To not use the REAGENT CONTROL function,
select the NOT IN USE option.
COUNT LIMIT sets the count limit for the asso-
ciated REAGENT CONTROL options (the
COUNT LIMIT indicates the maximum number of
cassettes/runs that may be processed before re-
agent replacement is necessary).
Touch the desired field to display the data entry
keypad. Enter the desired value, then touch the
ENTER key to close the keypad.
The factory default settings for COUNT LIMIT are
1000 cassettes and 25 runs.

Table of Contents

Other manuals for Sakura Tissue-Tek Xpress x120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Tissue-Tek Xpress x120 and is the answer not in the manual?

Sakura Tissue-Tek Xpress x120 Specifications

General IconGeneral
BrandSakura
ModelTissue-Tek Xpress x120
CategoryComputer Hardware
LanguageEnglish