Revised 10/02/2012 3.3
• UPS – indicates whether or not an external Unin-
terruptible Power Supply (UPS) is connected to
the instrument.
Touch the appropriate option field, either IN USE
or NOT IN USE.
The factory default setting for UPS is IN USE.
• OUTPUT –To determine method to export vari-
ous data that Tissue-Tek Xpress x120 system
recorded.
Touch the desired option field, either PRINT or
FILEOUTPUT. The factory default setting for
OUTPUT is FILE OUTPUT. (See “To Print Sys-
tem Setting” on page 3.4 for more information).
• EDIT ITEM – indicates whether or not the USER
DEFINED FILE option is enabled (see “Managing
the User Defined File” on page 3.6 for more in-
formation).
Touch the appropriate option field, either IN USE
or NOT IN USE.
The factory default setting for EDIT ITEM is NOT
IN USE.
• FILE NAME – indicates the file name for the
USER DEFINED FILE.
The factory default setting for FILE NAME is
SampleItem.
3. Touch the NEXT button to display the SYSTEM
SETUP screen, page 2 (Figure 3-C). Make changes
to the following system settings/parameters on
SYSTEM SETUP screen, page 2, as desired:
Figure 3-C
• SOUND VOLUME - sets the volume for “key
clicks” and the audible “end of process” and
“alarm” indications.
Touch the desired option field, either HIGH (au-
dible within 100 feet of instrument), MIDDLE (au-
dible within 50 feet of instrument) or LOW (audi-
ble within 10 feet of instrument).
The factory default setting for SOUND VOLUME
is MIDDLE.
NOTE: Touch the TEST CLICK, TEST ALARM, or
TEST END OF PROCESS button to test the selected
SOUND VOLUME.
• SOUND SELECTION – allows for selection of
the desired “end of process” and “alarm” sounds
from one of seven available options.
Touch the desired field to display a drop-down
menu of available options, then select the de-
sired option.
The factory default settings for SOUND
SELECTION are ENDOFPROC1 and ALARM1.
NOTE: Touch the TEST ALARM or TEST END OF
PROCESS button to test the selected SOUND
SELECTION.
• END OF PROCESS SOUND INTERVAL – se-
lects the interval at which the “end of process”
sound is repeated.
Touch the desired option field, either
CONTINUOUS, INTERMITTENT, or SINGLE.
The factory default setting for END OF PROCESS
SOUND INTERVAL is CONTINUOUS.
• LANGUAGE – selects the display language for
the instrument display screens.
Touch the desired field to display a drop-down
menu of available options, then select the de-
sired option.
The factory default setting for LANGUAGE is
ENGLISH.
• DATE FORMAT – sets the format for the date
display shown at the top of all instrument display
screens.
Touch the desired option field, either YMD, MDY,
or DMY.
The factory default setting for DATE FORMAT is
MDY.