Revised 10/02/2012 1.1
Section 1
INTRODUCTION
Intended Use
The Tissue-Tek
®
Xpress
®
x120 is designed for the
purpose of processing human and animal tissue speci-
mens.
The instrument, as part of the histopathology process,
is intended to facilitate the in vitro examination of human
and animal tissue for morphology changes by a
pathologist.
General Description
The Tissue-Tek
®
Xpress
®
x120 Continuous Rapid Tis-
sue Processor is a self-contained tissue processor
employing microwave and vacuum infiltration techniques.
The instrument performs automatic rapid processing
(fixation, dehydration, clearing, and paraffin impregnation)
of tissue specimens in preparation for histological study
and examination. The instrument incorporates individual
processing stations, scheduling software, and an auto-
mated transfer system that provides for continuous
processing of multiple sample baskets. Processing time
for a single basket is approximately 65 minutes, with a
maximum throughput (during continuous processing) of
120 specimens per hour.
NOTE: Maximum throughput is based on processing 40
cassettes per basket, where processing time for all
processes is equal (15 minutes). Reduction in the number
of cassettes processed per basket, or an increase in
processing time will lower throughput.
In order to achieve continuous processing, the instru-
ment utilizes a Transfer System that sequentially moves
baskets containing tissue specimens from a Loading
Station, through the four processing retorts to an Unload-
ing Station for removal of the basket. The instrument
employs scheduling software to efficiently process and
transfer sample baskets.
The Tissue-Tek Xpress x120 instrument has the follow-
ing features:
Instrument can be configured for either Manual or
Automatic start-up
Compatible specimen types include Formalin fixed
tissue, tissue fixed in Tissue-Tek Xpress Molecular
Fixative
Supports use of standard cassettes and Tissue-Tek
®
Paraform
®
cassettes (both microwavable); compatible
cassettes include:
Standard cassette (Tissue-Tek
®
Brand Uni-Cassette
®
Cassettes)
Biopsy cassette (Tissue-Tek
®
Brand Uni-Cassette
®
Biopsy Cassettes)
Tissue-Tek
®
Brand Paraform
®
Standard cassette
Tissue-Tek
®
Brand Paraform
®
Biopsy cassette
Tissue-Tek
®
Brand Paraform
®
Biopsy 13 x13 cassette
Tissue-Tek
®
Brand Paraform
®
Biopsy Core cassette
Tissue-Tek
®
Brand Paraform
®
Biopsy Shaved cas-
sette
Tissue-Tek
®
Brand Paraform
®
Orientation cassette
Standard cylindrical basket with lid facilitates
processing of up to 40 cassettes per basket; upper
loading limit of basket is marked to aid in positioning
cassettes
Tissue-Tek
®
Paraform
®
magazines are compatible for
use with the instrument (for use in conjunction with
the Tissue-Tek
®
AutoTEC
®
Automated Embedding
System
)
. Each Tissue-Tek Paraform magazine holds
up to 20 Paraform
®
cassettes. Two magazines linked
by a handle can be placed in the Loading Station of
the Tissue-Tek Xpress x120.
Two user-selectable automated processing programs
based on specimen thickness, Standard Program (15
minutes per retort) or Extended Program (30 minutes
per retort).
Maximum throughput of 120 cassettes per hour
(under Standard Program, where each basket or pair
of magazines processed contains 40 cassettes)
Integral fume control system to prevent fumes from
escaping the instrument

Table of Contents

Other manuals for Sakura Tissue-Tek Xpress x120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Tissue-Tek Xpress x120 and is the answer not in the manual?

Sakura Tissue-Tek Xpress x120 Specifications

General IconGeneral
BrandSakura
ModelTissue-Tek Xpress x120
CategoryComputer Hardware
LanguageEnglish