EasyManuals Logo

Sakura Tissue-Tek Xpress x120 User Manual

Sakura Tissue-Tek Xpress x120
82 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #72 background imageLoading...
Page #72 background image
TROUBLESHOOTING
8.2 Revised 10/02/2012
PROBLEM PROBABLE CAUSE REMEDY
Warm-up or shutdown does not
begin according to Auto
Ready/Shutdownsetting.
1. Instrument is in Manual Ready/Shutdownmode.
2. NOT IN USEis set for the day in READY / SHUTDOWN
TIMESscreen.
3. The day is specified as holidayin “DAYS OFF” screen.
4. Date setting is not correct.
5. ON TIMEset for the day in READY/SHUTDOWN
TIMESscreen is too late.
1. Change setting to Auto Ready/Shutdown
mode.
2. Set ON TIMEand OFF TIME”.
3. Cancel holiday setting.
4. Correct date setting.
5. Set earlier ON TIME”.
Shutdown begins unexpectedly.
1. Instrument is in Auto Ready/ Shutdownmode.
2. OFF TIMEset for the day in READY/SHUTDOWN
TIMESscreen is too early.
3. Date setting is not correct.
1. Change setting to Manual Ready/Shutdown mode.
2. Set later OFF TIME”.
3. Correct date setting.
Automatic warm-up operation
ended approximately 30
minutes earlier than set time.
1. Since automatic warm-up begins one hour before ON
TIME” (in consideration of various environmental tempera-
tures), warm-up ends earlier than set time.
1. Use instrument as it is, or set later ON
TIME”.
Shutdown does not begin
according to Ready/Shutdown
Times setting (with instrument in
“Auto Ready/Shutdown” mode).
1. One or more baskets are currently being processed.
2. One or more baskets are unloading station.
3. Instrument was started manually following an automatic
shutdown (automatic shutdown will not begin until next
OFF TIMEoccurs when the instrument has been started
manually following automatic shutdown).
4. OFF TIMEsetting is not correct.
1. Continue processing; remove basket(s) from
unloading station when complete.
2. Remove basket(s) from unloading station.
3. Shut the system down manually.
4. Set correct OFF TIME”.
System started automatically
and then shutdown began
automatically after a time.
1. Power failure occurred. 1. No action is needed.
Monitored Errors
Certain components, operating parameters and system
conditions are monitored while the system is powered on,
and during operation, to ensure the instrument is in
operational condition. When a monitored component or
parameter falls outside the established limits or operating
criteria, an alarm sounds continuously until the error is
cleared, and an error number or warning message is
shown on the Tissue-Tek Xpress x120 display. To cancel
(turn off) the alarm, press the EXIT button on the warning
message.
Each error that results in the display of an error number
generates an entry in the system Error Log. The Error Log
stores the 100 most recent errors, arranged by date, and
provides a description of each error (see “Viewing the
Error Log” on page 4.18 for details).

Table of Contents

Other manuals for Sakura Tissue-Tek Xpress x120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Tissue-Tek Xpress x120 and is the answer not in the manual?

Sakura Tissue-Tek Xpress x120 Specifications

General IconGeneral
BrandSakura
ModelTissue-Tek Xpress x120
CategoryComputer Hardware
LanguageEnglish