APPLICATIONS MANUAL
A.6 Revised 10/02/2012
Do’s and Don’ts
DO leave breast or fatty tissue in Pre-Processing
Solution 30 to 60 minutes to firm tissue prior to trim-
ming.
DO NOT use blue sponges for processing of tissue.
DO use embedding bags, lens paper, etc.
DO NOT process tissue for RNA studies with
Formalin-fixed tissue or solutions that have been
used for Formalin-fixed tissue.
DO run RNA tissue after the processor is cleaned and
solutions are all changed.
DO NOT process tissue with O.C.T. compound (used
for Cryotomy) residue in the Tissue-Tek Xpress x120.
DO remove all O.C.T. compound (used for Cryotomy)
from specimens by rinsing with water prior to placing
in the Tissue-Tek Xpress x120.
DO NOT process tissue thicker than 2 mm in
thickness on the Standard program.
DO process tissue that is more than 3 mm in
thickness using the Extended program.
DO NOT allow the faced tissue blocks to soak too
long in water.
DO face the block and allow the block to soak/or cool
only for a short period of time.
DO NOT use metal cassette lids.
DO use plastic lids and remove metal staples and
sutures from tissue prior to placement in the cassette.
DO NOT top off reagents or rotate reagents.
DO change the reagents and solution after 1,500
cassettes or every 3 days, whichever occurs first.
DO seal blocks after Microtomy with a thin layer of
paraffin if block will later be used for molecular stud-
ies
Disposal of Tissue-Tek Xpress
x120 Processing Solutions
Refer to MSDS sheets. Disposal should be in accord-
ance with Local, State and Federal regulations.

Table of Contents

Other manuals for Sakura Tissue-Tek Xpress x120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Tissue-Tek Xpress x120 and is the answer not in the manual?

Sakura Tissue-Tek Xpress x120 Specifications

General IconGeneral
BrandSakura
ModelTissue-Tek Xpress x120
CategoryComputer Hardware
LanguageEnglish