Revised 10/02/2012 2.1
Section 2
INSTALLATION
Turning Instrument Power On or Off
The main power to the entire instrument is turned on
and off by the power switch
.1
. (Figure 2-A), located on the
right side panel. In the up position the switch turns the
power on and in the down position the switch turns the
power off.
Figure 2-A
Turning the Power Off (for extended nonuse)
The Tissue-Tek
®
Xpress
®
x120 instrument has been
designed to run continuously. However, if it ever becomes
necessary to turn the unit off for extended periods (weeks
or months), it is recommended that all reagents be drained
from the instrument and the instrument processing
chamber be thoroughly cleaned and disinfected. The
loading station, unloading station, and retort access doors
should then be left open until the interior of the instrument
is completely dry. Once dry, close the access doors to keep
the interior free of dust and debris.
General Information
This section provides information on determining a
location for, and installing, the Tissue-Tek
®
Xpress
®
x120
instrument. Installation should be performed by a qualified
instrument service technician only. The Tissue-Tek Xpress
x120 instrument must be installed correctly to ensure
proper operation and service.
Read this operating manual carefully before attempting
to operate the Tissue-Tek Xpress x120 instrument. Follow
all instructions carefully.
CAUTION: The Tissue-Tek Xpress x120 is a precision
instrument and must be handled accordingly. Rough
handling or dropping will disturb or damage internal
components. Always handle the instrument with care.
Environmental Factors
As with all sensitive electronic equipment, avoid pro-
longed exposure to excessive temperature and humidity.
Temperature and humidity should be held relatively
constant to obtain the highest degree of operating stability.
The ambient operating temperature range of the instrument
is 15° to 35°C (59° to 95°F). The ambient operating
humidity range is 30% to 85% relative humidity, non-
condensing.
Locate the instrument in a well-ventilated area, avoid-
ing exposure to corrosive vapors, direct air currents, or
temperature extremes. Avoid proximity to direct sunlight,
open windows, sinks, ovens, open flames, hot plates,
radiators, and dry ice baths. Locate the instrument away
from any equipment that consumes a high voltage or large
current, including large refrigerators and ovens. Since the
instrument is very heavy, the floor must be solid and level.
Be sure the instrument is located near a power source
that meets the electrical requirements (voltage and
amperage) specified on the rating label located on the rear
of the instrument. The power receptacle must be grounded
and should be a clean, noise-free, dedicated line.
Ensure that the selected installation site provides suffi-
cient clearance to allow for proper operation if the instru-
ment, and to provide adequate ventilation for the conden-
ser, vacuum pump, microwave units, fans, and pumps.

Table of Contents

Other manuals for Sakura Tissue-Tek Xpress x120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Tissue-Tek Xpress x120 and is the answer not in the manual?

Sakura Tissue-Tek Xpress x120 Specifications

General IconGeneral
BrandSakura
ModelTissue-Tek Xpress x120
CategoryComputer Hardware
LanguageEnglish