The calibration procedure is as follows:
Connect the short load.
Load status is indicated by the color of the
voltage and current detector indicators.
Green denotes a load good status.
Press the appropriate button to continue, or
to exit.
The calibration progress will be shown in
the display bar.
The calibration is completed.
The calibration coefficients are shown at
the bottom of the screen, for information.
Press the appropiate button to apply the
new calibration settings, or to cancel them.
Press the appropiate button to go to the
optional Detector Test, or to exit.