Operating Instructions Stainless Steel In-Line Filter Holder Type 16274 17
Getting Started
t Remove the perforated sheet from the lower part of the device.
t Remove the three device legs from the lower part of the device.
t Remove the valve from the lower part of the device.
t Remove the tube nozzle from the lower part of the device.
5.1.2 Cleaning the Device
Requirement
The device is dismantled.
Procedure
t NOTICE Corrosion or damage to the device due to unsuitable cleaning
agents! Clean the device and all disassembled device components with
warm water, a soft brush and gentle, commercially available laboratory
cleaning agents. The cleaning agent must be suitable for metal, glass,
and plastic.
t Rinse the device and all device components with hot water after cleaning
them.
t Rinse the device and all device components with distilled water.
t NOTICE Fiber residues on the device due to the use of cloths! Dry the
device and all device components in the compressed-air flow or leave to
air dry, do not dry with cloths.
5.1.3 Reassembling the Device
Procedure
t Assemble the device (see Chapter “4.4 Mounting the Device”,
page 14).