Operating Instructions Stainless Steel In-Line Filter Holder Type 16274 27
Technical Data
12 Technical Data
12.1 General Data
12.2 Dimensions and Weight
Unit Value
Membrane filter which fits the diameter mm 142
Prefilter which fits the diameter mm 130
Filtration area cm² 130
Capacity ml 2000
Weight kg 12
12.3 Ambient Conditions at the Installation Site
Unit Value
No heat from heating systems or direct sunlight
12.4 Operating Conditions
Unit Value
Maximum operating pressure bar 7
Maximum operating temperature °C 80
Maximum temperature during autoclaving for a maximum of
30 minutes at a pressure of 2 bar
°C 134
Maximum temperature for sterilization with dry heat, depres-
surized, depending on the filter used
°C 180
12.5 Device Connections
Inlet and outlet M 12 x 1 internal thread