37
– To store the setting, press and hold the ) key (> 2 sec).
– Select menu item1-19-2 (key sequence: k, )).
– Enter the resolution (lowest scale interval d) (e.g., 0.01 kg). This value has to
match that stored under menu item 11-4-1 (or 11-5-1 or 11-6-1).
– To store the setting, press and hold the ) key (> 2 sec).
– Select menu item1-19-3 (key sequence: k, )).
Enter the sensitivity of the load cell in mV/V (usually listed in the load cell
specification sheets). If a weighing platform consists of more than one load cell,
enter here the sensitivity of load cell 1.
Example: The sensitivity of the load cell is 1.9440 mV/V.
– To store the setting, press and hold the ) key (> 2 sec).
Note: If a weighing platform consists of more than one load cell, enter the sensitivity
values for the other cells (up to 4) under menu items 1-19-4 (for load cell 2)
through 1-19-6 (for load cell 4) and confirm each value (k ) keys).
If the platform has less than four load cells, enter “0.000” under the unused menu
items.
display_schrittw_001Ib.eps
display_wze19440.eps