48  Operating Instructions Combics Indicators
  Operation
(
  1.)  Zero the scale.
J
  2.)   Start calibration 
(e.g., when adjustment 
prompt flashes WP).
   cal.Ext. is displayed for 
two seconds.
   You are prompted to place 
the required weight on the 
platform (e.g., 10,000 g).
  3.)    Place the calibration/-
adjustment weight on the 
weighing platform.
   The difference between 
the measured value and the 
true weight of the sample 
will be displayed with plus/
minus signs.
Ext.     calibration A printout will be 
Nom.  +      10000 g generated if the process  
Diff. +          1 g    is cancelled using the (
J  4.)   Activate calibration adjust-
ment (press the ( key to 
cancel).
   The adjustment weight  
is displayed once adjust-
ment is finished.
--------------------  A GMP-compliant 
24.02.2012     10:15  printout is generated.
Type          CAISL2
Ser.no.     12345678
Vers.    1.0103.11.2
BVers.      01-26-02
-------------------- 
Ext.     calibration 
Tar   +      10000 g 
Diff. +          1 g
Ext.      adjustment 
Diff. +          0 g
--------------------
24.02.2012     10:15
Name:
--------------------
Example:
External calibration and manual adjustment with default weights (weighing parameters: factory settings)
Example 2
External calibration and manual adjustment with freely configurable adjustment weight (within the limits of 1/3 max. capacity to max. capacity)
(
  1.)  Set scale to zero.
J
  2.)  Start calibration (e.g.,  
    when adjustment prompt  
    flashes WP).
  “cal.ext.” is displayed 
   for two seconds.
You are prompted to place the 
calibration/adjustment weight on 
the platform (e.g., 10,000 g).
5  +  )  3.) Enter the desired CAL weight 
above the number block and 
confirm with “)” (5,000 g here). 
If the weight is too heavy or 
too light, an error message will 
appear.