85 Operating Instructions Combics Indicators
Totalizing s (Combics 2)
With this application, you can add weights to the totalizing memory. In addition to
weight values, the number of separate values added to memory is also saved
(APPLIC.3 menu).
Features You can weigh up to 999 items.
Save values automatically: Simultaneous saving of net values and calculated
values (if available).
Setting: Applic./APPLIC.3 /TOTALIZ menu item 3.16.
Save weight values and calculated values from either Application 1 (for example,
Counting, Weighing in Percent) or Application 2 (Checkweighing).
Setting: Applic./APPLIC.3 /TOTALIZ menu item 3.22.
Current transaction number displayed in the number display (indicating the
transactions already added)
Weighing in up to a defi ned target, with the totalization memory content +
current weight displayed in the text lines
Save weight values manually or automatically
Accurate calculation of total of weight values from two weighing platforms
Activate Info mode via the I key
Automatic printout when value saved
Automatic taring of container weight.
Setting: Applic. / aut.tare, menu item 3.7
Exit application, delete parameters
The value of the totalizing memory remains valid until deleted by pressing the
c key.
You can assign different functions to the c key for deleting applications.
When you clear applications, you can delete either the data stored for all
applications or just selected data stored for the active application.
Setting: Applic./CLER.CF menu item 3.24
Tare function:
If you store a tare (weight value) by pressing the ) key, you can later enter a tare
value manually. The tare value you enter is added to the stored tare value.
Setting: Applic. / TARE.FNC menu item 3.25.1 (factory setting)
A tare value entered manually overwrites a stored tare value (weight value). If you
enter a tare value manually, a tare value (weight value) stored later overwrites the
manually entered value.
Setting: Applic. / TARE.FNC menu item 3.25.2
Restore factory default settings: Applic./Reset menu item 9.1.
The Combics has a totalizing memory for adding individual net and gross values.
You can save weight values in totalizing memory manually or automatically.
Setting: Applic./APPLIC.3 /TOTALIZ menu item 3.16.
Save value manually by pressing the O key.
The value taken from the active platform is added to the value already saved
in totalization memory and the transaction counter value is increased by one.
When a value is added manually, the program does not check whether the
platform has been unloaded since the last time the O key was pressed.
Operating Instructions Combics Indicators 85
Operation

Table of Contents

Other manuals for Sartorius combics 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius combics 2 and is the answer not in the manual?

Sartorius combics 2 Specifications

General IconGeneral
BrandSartorius
Modelcombics 2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals