Entering the reference piece weight
The reference piece weight (i.e., the weight of one piece) can be entered using the
keypad and save with the O key.
Reading the reference piece weight
The reference piece weight can be read using a barcode scanner.
h
The entered value remains active until deleted by pressing the c key or until
overwritten by a new value. It remains saved after the scale is switched off.
Preparation t Open the APPLIC. / APPLIC.1 / COUNT. menu.
Available parameter settings * = Factory setting
MIN.INIT Minimum load for initialization 3.6
1 digit 1 digit* 3.6.1
2 digit 2 digits 3.6.2
5 digit 5 digits 3.6.3
10 digT. 10 digits 3.6.4
20 digT. 20 digits 3.6.5
50 digT. 50 digits 3.6.6
100 diG. 100 digits 3.6.7
200 diG. 200 digits 3.6.8
500 diG. 500 digits 3.6.9
1000 dI. 1000 digits 3.6.10
RESOLUT Resolution for calculation of reference value 3.9.
DISP.ACC Display accuracy* 3.9.1
10 FOLD Display accuracy + 1 decimal place 3.9.2
100FOLD Display accuracy + 2 decimal places 3.9.3
SAVE WT. Parameter for saving weight values 3.11
STABIL. With stability* 3.11.1
ACC.STAB With increased stability 3.11.2
REF.UPDT Reference sample updating . 3.12.
OFF Off 3.12.1
automat Automatic* 3.12.3
REF.WP Reference weighing instrument 3.13.
NO WP No weighing platform selected 3.13.1*
wp 1 Weighing platform WP1 3.13. 2
wp 2 Weighing platform WP2 3.13. 3
) t To save the setting, press the ) key.
( ( ... t To exit setup: Press the ( key several times.
Minimum load for initialization You can set the minimum load here, i.e. the load that must be placed on the
weighing platform in order to carry out the application. If the load on the platform
is too light, the following will occur:
Error code Inf 29 appears,
The weighing platform is not initialized,
The preset reference sample quantity is saved.
Setting: Applic./Applic.1/Count./Min.init menu item 3.6.
The minimum load can be set in 10 steps from 1 to 1000 digits (see available
parameters). The “digits" here refer to the scale intervals for the connected weighing
platform. If the interval of the connected platform is 1 g, for example, and 1000 digits
are required, you must place at least 1000 g (= 1000 intervals = 1000 digits) on the
weighing platform for initialization.
Operating Instructions Combics Indicators 55
Operation

Table of Contents

Other manuals for Sartorius combics 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius combics 2 and is the answer not in the manual?

Sartorius combics 2 Specifications

General IconGeneral
BrandSartorius
Modelcombics 2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals