Entering the reference percentage value
The reference weight for 100% is entered using the keypad and the O key is
pressed to initialize the application.
Reading the reference percentage value
The reference weight can be read using a barcode scanner.
h
The entered value remains active until deleted by pressing the c key or until
overwritten by a new value. It remains saved after the scale is switched off.
Preparation t Open the APPLIC. / APPLIC.1 / PERCENT menu
Available parameter settings * = Factory setting
MIN.INIT Minimum load for initialization 3.6
1 digit 1 digit* 3.6.1
2 digit 2 digits 3.6.2
5 digit 5 digits 3.6.3
10 digT. 10 digits 3.6.4
20 digT. 20 digits 3.6.5
50 digT. 50 digits 3.6.6
100 diG. 100 digits 3.6.7
200 diG. 200 digits 3.6.8
500 diG. 500 digits 3.6.9
1000 dI. 1000 digits 3.6.10
RESOLUT Resolution for calculation of reference value 3.9.
DISP.ACC Display accuracy 3.9.1*
10 FOLD Display accuracy + 1 decimal place 3.9.2
100FOLD Display accuracy + 2 decimal places 3.9.3
DEC.PLCS Decimal places in displayed result 3.10
NONE None 3.10.1*
1 DEC.PL 1 decimal place 3.10.2
2 DEC.PL 2 decimal places 3.10.3
3 DEC.PL 3 decimal places 3.10.4
SAVE WT. Parameter for saving weight values 3.11
STABIL. With stability 3.11.1*
ACC.STAB With increased stability 3.11.2
REF.WP Reference weighing instrument 3.13.
NO WP No weighing platform selected 3.13.1*
wp 1 Weighing platform WP1 3.13.2
wp 2 Weighing platform WP2 3.13.3
CALC.DIS Display of calculated value 3.15.
RESID.QT Residual quantity 3.15.1*
LOSS Loss 3.15.2
) t To save the setting, press the ) key.
( ( ... t To exit setup: Press the ( key several times.
Minimum load for initialization You can set the minimum load here, i.e. the load that must be placed on the
weighing platform in order to carry out the application. If the load on the platform
is too light, the following will occur:
Error code Inf 29 appears,
The weighing platform is not initialized,
The preset reference sample quantity is saved.
68 Operating Instructions Combics Indicators
Operation

Table of Contents

Other manuals for Sartorius combics 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius combics 2 and is the answer not in the manual?

Sartorius combics 2 Specifications

General IconGeneral
BrandSartorius
Modelcombics 2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals