A tare value entered manually overwrites a stored tare value (weight value). If you 
enter a tare value manually, a tare value (weight value) stored later overwrites the 
manually entered value.  
Setting: Applic. / TARE.FNC menu item 3.25.2 
Restore factory default settings: Applic./Reset menu item 9.1.
  Target   Checkweighing entails comparing the current weight value to a defined target. You 
can enter the value for this target using the keypad, or by saving the weight value 
indicated. You can also define upper and lower tolerance limits based on this target. 
You can do this by:
–   Entering absolute values using the keypad or placing the desired amount of 
weight on the platform and saving the value 
or
–   by entering each value as a percentage deviation of the target weight
or
–   by entering each value as an asymmetrical percentage deviation of the target 
weight that is selected via the keypad or using the r key 
or
–   by entering a relative weight deviation from the target weight via the keypad.
The value remains valid until deleted by pressing the c key or until overwritten by 
a new value. It remains saved after the scale is switched off. 
  Preparation  t  Open the APPLIC. / APPLIC.2 / CHECK.WG menu.
  Available parameter settings   * = Factory setting
MIN.INIT Minimum load for initialization   3.5
  1 digit  1 digit  3.5.1*
  2 digit  2 digits  3.5.2
  5 digit  5 digits  3.5.3
  10 digT.  10 digits  3.5.4
  20 digT.  20 digits  3.5.5
  50 digT.  50 digits  3.5.6
  100 diG.  100 digits  3.5.7
  200 diG.  200 digits  3.5.8
  500 diG.  500 digits  3.5.9
  1000 dI.  1000 digits  3.5.10
Aut.strt  Automatic start of applications when you switch on  
the device with most recently saved application data        3.8
  automat  Automatic (on)  3.8.1*
  manual  Manual (off)  3.8.2
 TARE.FNC Tare function    3.25
  normal   Can add a preset tare if tare value is avail- 
able; however no tare function possible  
3.25.1*
  special   When a preset tare is entered, the tare 
value is deleted; however, tare function 
activation is possible  3.25.2
CHECK.RG  Checkweighing range   4.2.
  30- 170%  30 to 170%  4.2.1*
  10-max.l  10% to infinity  4.2.2
  Operating Instructions Combics Indicators  73
  Operation