Changing the Resolution
Purpose
To change the resolution of the
weighing result. This enables quicker
weighing at a lower resolution.
Features
Weights can be displayed with a
reduced resolution.
Once this function is selected, the dis-
play will appear as usual, e.g.,
» ... d=0.01 mg« will be shown. To tog-
gle to the 4-digit range:
press the
d*10 key.
In the metrological line of the display,
the readout will toggle accordingly to
» ... d=0.1 mg«. Afterwards, the soft key
should be labeled with
d/10.
In general, the balance is tared each
time the number of decimal digits dis-
played is changed. This ensures that the
unit is precisely tared in accordance
with the regulations governing legal
metrology each time the readout is
toggled.
This function can be assigned to either
the fourth or fifth soft key from the
right (F4 or F5).
The soft keys are labeled as follows:
during normal resolution:
d*10
during reduced resolution:
d/10
The display cannot be zeroed if any
object is loaded on the weighing pan. In
this case, the error message
ERR 08 oO zero range will
appear.
Factory settings of the parameters
No parameters can be set.
Preparation
§ Turn on the balance: press e
> The Sartorius logo is displayed
§ Select Extra function (F4) or Extra function (F5) in the Setup menu: press s
§ Select
Application parameters: press the q soft key twice,
then O soft key once
§ To select
Extra function (F4) or Extra function (F5):
press the q soft key repeatedly, then O soft key once
§ Select
Change Resolution
§ Confirm the Change Resolution function.
See also: “Application Parameters (Overview)” in the chapter entitled “Configuring the
Balance”
§ Store the settings and exit the setup menu: press the
oo soft key.