EasyManua.ls Logo

Sartorius ME5 - Device Information; Factory Settings

Sartorius ME5
169 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Purpose
This menu item enables you to have
information displayed about the specific
balance (“device”).
Display Device Information on
ME215/235/254/414/415/614
§ Select the Setup menu:
press the s key
> “SETUP” is displayed:
§ Select “Info”:
Repeatedly press the
q soft key, then
press the
O soft key
§ Select “Device information”:
press the
O soft key
> Device information is displayed:
Display Device Information
on ME5 or SE2
§ Select “Device Information”:
Press the j key
> Device information is displayed
oo o q
SETUP INFO
Version no: 01-41-05
Wgh.sys. ver. #: 00-21-09
Draft sh. ver.#: 05-01-03
Model: ME215S
Serial no: 91205355
§ Print device information:
Press the p/r key
> Printout (Example)
--------------------
23.12.2001 13:02
Model ME215S
Ser. no. 91205355
Ver. no. 01-41-05
(Version of the operating program)
ID BECKER123
(User ID)
--------------------
L ID LOT 23
(Lot ID)
SETUP
INFO
--------------------
Version no.:
01-41-05
(Version of the operating program)
Wgh. sys. ver. #:
00-21-09
(Version no. of the weighing cell)
Dft. shield v. no.:
05-01-03
(Program version no. of the
draft shield)
Model:
ME215S
Serial no.:
91205355
Next mainten.:
01.01.2003
Service phone:
00495513080
SQmin:
0.0300 g
--------------------
§ Return to SETUP overview:
press the
o soft key
§ Exit Setup menu:
Press the
oo soft key
> Original settings are restored
Factory Settings
Each parameter category has a factory
setting. In the Setup menu, you can
restore all factory settings by confirming
the selection
YES.
The following settings are not restored:
Language
Password
Display contrast
Time (clock)
38
Device Information

Table of Contents

Related product manuals