Purpose
With this application, you can have
weight values and calculated values
totalized and statistically evaluated.
The values determined for the
evaluation are:
average (mean value)
standard deviation
variation coefficient
sum of all values
lowest value (minimum)
highest value (maximum)
difference between the minimum and
the maximum
You can use the statistics application in
combination with any program chosen
from Application 1 (such as counting,
weighing in percent) and one from
Application 2 (checkweighing, time-
controlled functions) as well as with the
extra functions.
Features
Storage of weight values and calculated
values
Simultaneous storage of net and
calculated values
Optional configuration in the Setup
menu for loading weight values and
calculated values either from Applica-
tion 1 (such as counting, weighing in
percent) or from Application 2 (check-
weighing, time-controlled functions)
Totalizing memory for up to 65,535
values
Simultaneous display in the text line of
the transaction counter and, e.g., the
current total
Optional configuration in the Setup
menu for having the balance tare auto-
matically after a value has been stored
in the totalizing memory
Manual input of the number of
individual weighing operations and
confirmation using the
nDef soft key
(target no. of operations nDef). Result
printed and memory cleared after
printout of nDef
Optional configuration in the Setup
menu to add the current weight, with
display accuracy, to the current total by
pressing the
M+ soft key and generate a
printout of the result
Optional configuration in the Setup
menu for stability-dependent storage
of the measured value: Balance/scale
functions, Stability range
Optional automatic storage of measured
values
Storage of measured value is indicated
by
Uu;
zz indicates that you can place a load
on the balance
Minimum load threshold for automatic
storage
Press the
M- soft key to delete the last
value added to the totalizing memory.
The transaction counter value is
reduced by one and a printout is
generated
Press the
MR: soft key for information
about number of transactions and the
current total. By configuring the Setup
menu, you can define whether the
information is displayed and printed, or
only printed, and whether the informa-
tion comprises an intermediate
or final evaluation
In the Info window you can use the
q,
l (o) soft keys to choose which value
will be displayed in the text line during
weighing
Printout of the final result depending
on the Application 1 or Application 2
parameters. Configure the Setup menu
to define which values are included on
the printout (printout of individual
components)
Press
MR for a printout of an inter-
mediate evaluation after each addition
or a final evaluation
A final evaluation is printed when the
statistics routine is ended by pressing
c, if no final evaluation was gener-
ated by pressing
MR
Optional configuration in the Setup
menu to clear the statistics memory and
reset the transaction counter by press-
ing c or after an evaluation is print-
ed out
Totalization data and transaction
counter data are stored in the non-
volatile memory
Continue totalization after turning the
balance off and back on