67
Features
You can store an equation and
configure the Setup menu to initialize
this program automatically with the
stored equation
The
o symbol is displayed to indicate
a calculated value; the equation
used is displayed in the text line
If no equation was entered, the weight
value is displayed
Toggle between the weight readout,
equation input and display of the
calculated result by pressing the
corresponding soft keys (or press
c to toggle between weight and
calculated value)
There are four operators
(+, –, *, /) and one factor (weight value)
available when you enter an equation
Max. equation length:
28 characters
Pressing c will delete either
the equation or the last character
entered, depending on the
configuration in the Setup menu
(Setup: Device parameters: Keys:
CF function for inputs: Delete last
character)
The calculated result is displayed
with the number of decimal places
configured in the Setup menu.
Not all decimal places are displayed
if the result is longer than the display
allows. If there are more digits before
the decimal point than the display can
show, an error message is displayed
The equation is stored in
non-volatile memory
Factory Settings of the Parameters
Decimal places in calculated result:
2 decimal places
Soft Key Functions
Equat. Toggle to equation
+ Enter an addition operator
in the equation
- Enter a subtraction operator
in the equation
* Enter a multiplication
operator in the equation
/ Enter a division operator
in the equation
Start Start calculation
Weigh Toggle to the weighing mode
Weight Enter a weight operand
in the equation
Printout for Calculation
The calculation result is printed.
Res + 693.88 o
Res: Result of calculation with equation
Purpose
With this application you can calculate
a weight value using an algebraic equation.
This can be used, for example, to deter-
mine the weight per unit area or “gsm”
weight (grams per square meter) of paper.
You can use this application in
combination with any program chosen
from Application 2 (checkweighing,
time-controlled functions) and one from
Application 3 (totalizing, formulation,
statistics) as well as with the extra
functions.
Calculation D