Example
Determining an Unknown Piece Count; Weighing-In the Preset Reference Sample Quantity
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Application 1: Counting: Average piece weight updating: Manual
Setup: Application-defined output: Auto print upon initialization: All values
Step Press key(s) (or follow instructions)  Display/Output
1. Delete previous values if necessary c
2. Prepare a container for the Place the empty container
parts to be counted on the balance
3. Tare the balance t
4. Place the reference sample  Place the displayed number
quantity on the balance  of parts in the container
(example: nRef = 10 pcs)
5. Determine the average piece  Start soft key
weight (number of decimal 
places displayed depends on 
the balance model)
(If you do not need a printout, select
this setting in the Setup menu)
6. If necessary, increase the number Place additional parts 
of parts and perform reference in the container
sample updating (example:
Update soft key
7 additional parts)
7. Weigh uncounted parts Place parts to be counted
in the container
8. If desired, print total piece count p/r
(here: 72 pcs)
63